|
mechanical
engineering design
2004
blood ties
An engineering team rethinks the key link in cardiac surgery, the heart-lung machine.
The heart is an amazing organ. in reality,
it's a double pump, with one side pushing blood to the lungs and
the other side pushing the oxygen-rich blood returning from the lungs
out to the rest of the body.
Because of this complexity, when something goes wrong with the heart,
fixing it is a complicated task. Stop the heart to make a repair and the
lungs stop, too. In order for the surgeon to operate on the heart, an
outside system must provide the pumping action to get blood through the
body, and it must assume the function of the lungs, removing carbon dioxide
from the blood and letting oxygen in.
During open-heart surgery, such as a bypass procedure, a machine takes
over the function of both heart and lungs. This heart-lung machine allows
the surgeon to carefully stop the heart, while the vital organs continue
to receive blood and oxygen. The surgeon can then set about the delicate
task of repairing the damaged heart.
 |
| The Synergy system uses a centrifugal
pump to move a patient's blood. |
Heart-lung machines, in use in one form or another since
the 1950s, have saved countless lives. But they can also introduce complications
of their own. And some of those complications can be life-threatening.
That's why a European engineering enterprise has developed a machine
designed to get around some of the most serious side effects of the heart-lung
machine.
The conventional heart-lung machine pumps blood through a single, disposable
circuit. The circuit consists of a large bucket, or reservoir, used to
hold extra blood volume; a peristaltic or roller pump that pushes blood
through the body; a heat exchanger to regulate the patient's temperature;
an oxygenator used to oxygenate the patient's blood and remove
carbon dioxide waste, and a filter to help remove particulates, or emboli,
from blood before sending it back to the patient.
Blood is drained by gravity from the patient, then pumped from the large
reservoir through the heat exchanger, oxygenator, and filter, then back
into the patient's body. All the components are attached to one
another using various lengths of PVC tubing that are also attached to
the patient. The nondisposable hardware portion of the heart lung machine
is called the pump console and directly manages this circuit.
The pump console regulates pump flow, monitors the patient's vital
parameters like blood pressure, temperature, and blood flow, and provides
real-time information on key physiological parameters like blood oxygen
saturation and levels of blood clotting.
Open-and-Shut Cases
This technology has improved significantly in terms of performance and
reliability over the past 25 years, permitting surgeons to make a formerly
risky procedure like open-heart surgery one of the most common procedures
for the treatment of heart problems. In 2000, 686,000 open-heart procedures
were performed in the United States alone, according to the American Heart
Association.
While mortality rates have improved greatly over the years, morbidity
and procedure-related postsurgery sickness are still possible complications.
The circuit must be filled, or "primed," with a physiological
saline solution before the procedure in order to remove the air from the
circuit and to avoid removing too much blood from the patient. Priming
can cause postoperative complications like low red blood cell counts due
to over-dilution from the saline solution, particularly in smaller patients,
who do not have much blood volume in the first place.
Another problem is agitating the blood by removing it from the body and
running it through a plastic circuit.
Blood reacts to being in contact with strange surfaces, and blood components
like clotting factors and platelets can become so agitated that the blood
itself can present challenging problems for the patient after surgery.
This condition is called systemic inflammatory response and can significantly
lengthen a stay in the intensive care unit. That translates into higher
costs for the hospital, up to $1,000 a day in the ICU, a particularly
important consideration in today's environment of cost-based managed
health care.
In an effort to mitigate some of the issues relating to on-pump procedures,
minimally invasive cardiac surgery, or MICS, doesn't stop the heart.
Instead, surgeons work on it while it is still beating.
While recent advances in this field have been promising, the clinical
results of recent multiyear studies, such as one published in The Annals
of Thoracic Surgery in 2001 by David A. Bull and his colleagues at
the University of Utah Health Sciences Center, indicate that no significant
advantages using this approach are making themselves obvious. The minimally
invasive approach can also be difficult technically for the surgeon, who
must stitch a new blood vessel onto the surface of the heart while it
is still beating.
 |
| A traditional heart-lung system
uses multiple machines to pump blood through a single, disposable
circuit. |
"While the MICS approach will definitely be the long-term future
of this market," said Matteo Glauber, chief cardiac surgeon of
the Ospedale Pasquinucci in Massa, Italy, "we must do something
now to address patient morbidity using the heart-lung machine, while the
industry develops easier and more effective techniques for beating-heart
surgery."
The engineering staff at Dideco SpA in Mirandola, Italy, responsible for
improving the company's heart-lung bypass system, took this as
a challenge. Dideco and its sister divisions, Cobe Cardiovascular in Arvada,
Colo., and Stckert Instruments of Munich, Germany, worked together
to develop a system based on a heart-lung machine, but with a different
approach they believe will lead to fewer post-op complications.
"We really had our work cut out for us," said Dideco engineering
director Ivo Panzani. "First, to decide precisely how to improve
what we have been doing successfully for 20 years, but also how to convince
our management to accept such a high-risk project. There were no guarantees
of its success, and in fact there still are no guarantees it will be successful.
But, we are convinced this approach is the future of our on-pump business."
Dideco's system, called Synergy, functions as a heart-lung machine,
but handles the draining of a patient's blood differently from
the standard device. Typically, the perfusionist, who manages the heart-lung
machine, uses the reservoir to drain the patient's blood. Synergy
eliminates this reservoir completely and effectively uses the patient
as the reservoir.
The Synergy device consists of the same type of components as in a heart-lung
machine, but places them in a single integrated device. The components
are a pump, a heat exchanger, a blood oxygenator, and a filter. In a traditional
heart-lung circuit, each of these parts is separate, but in the Synergy
system they are combined into one assembly.
Drawn From the Heart
The Synergy system is based on a centrifugal pump that can be used to
actively suck the blood directly from the patient instead of draining
blood by gravity. Synergy is small30x20x20 cmand so is its
control console, with a footprint of 1.3 square meters, compared with
3 square meters for the traditional console.
As a result, the device can be placed closer to the patient. Therefore,
less tubing runs to and from the patient. Priming volume also drops, from
1,500 or more cubic centimeters for a standard heart-lung system down
to 900 cm3.
According to Roel de Vroege, a professor of clinical perfusion at the
Free University Hospital at Amsterdam in the Netherlands, "Since
the volume is so much smaller, you can use some of the patient's blood
to push the saline prime out of the circuit and reduce the prime even
more. We see almost no dilution of the patient's blood using this technique,
which is possible also with traditional circuits, but the effect is nowhere
as dramatic."
Rene Huybregts, a cardiac surgeon at the Free University Hospital and
one of those responsible for the first clinical evaluations of Synergy
in Europe, sees the reduction of blood dilution as critical to improving
the recovery of the patient. "Lower prime means that we should not
need to give transfusions in the intensive care unit, and everyone knows
in this field that lower transfusion rates are better for patients in
the long run."
Additionally, the Synergy system is totally coated with a phosphorylcholine
biocompatible coating intended to mimic the endothelium, the inside surface
of the body's blood vessels. The coating fools the blood into thinking
it is still inside the body, even though it is circulating through plastic
tubing. Since the blood still thinks it is in the body, the level of activation
of the blood clotting mechanisms and systemic inflammatory response mechanisms
in the blood are greatly reduced.
Additionally, a cell-washing machine called an autotransfusion machine,
or cell saver, is used with this system to suction any blood released
into the chest during surgery. Recent studies have shown that the spilled
blood coming from the chest has been highly "activated." The
use of the autotransfusion machine means that the highly activated blood
is cleaned and effectively deactivated before being returned to the patient.
Autotransfusion is currently not used routinely with a traditional system,
and a cell saver is used perhaps by only 30 percent of surgical centers
in the United States and Europe. Typically, blood is filtered and debubbled
in the reservoir of a traditional system. Because the Synergy system doesn't
have a reservoir, it must use an autotransfusion system.
The Synergy system uses a centrifugal pump from Cobe Cardio-
vascular to handle the pumping duties.
"The Synergy concept doesn't work without this pump," said Dideco's
Panzani. "We typically use a peristaltic roller pump in traditional
heart-lung machines, but such a system in our Synergy approach is a little
dangerous because the excess suction could damage the heart."
According to Panzani, the centrifugal pump reduces risk that can arise
when a surgeon moves the heart around to work on it. The movement sometimes
causes the stopped heart to collapse around the cannula. When the blood
flow is interrupted, a centrifugal pump will stop trying to pull fluid.
But a peristaltic roller pump, which is used to push blood in a conventional
heart-lung machine, is a positive displacement pump. If it were used to
draw blood and the heart were to collapse around the cannula, the pump
would not stop drawing and could damage heart tissue.
Dideco faced other important design considerations as well. Typically
with an oxygenator design, the focus is on using program simulations like
computational fluid dynamics to detect and eliminate areas of recirculation
and stagnation within the device. Such areas allow clots to form. If a
clot ends up in the patient, it can create an embolism that causes a stroke.
 |
| The Synergy system uses an oxygenator
to manage blood gases and a centrifugal pump to handle blood movement,
and places both close to the patient. |
In the case of Synergy, however, the initial problem was that the small
size of the device caused excess shear, to which some components of the
blood are extremely sensitive. Red blood cells exposed to excess shear
will tear open, releasing free hemoglobin into the bloodstream that can
potentially cause kidney malfunction. Platelets under excess shear become
extremely agitated, heightening the systemic inflammatory response. Even
with a biocompatible coating, excess shear can have the same effect postoperatively
on the patient as an uncoated circuit.
Dideco used CFD simulation to define and eliminate areas of excess shear,
particularly in the centrifugal pump and in areas where there were sudden
velocity changes within the device. The results were validated by laboratory
testing, focusing on levels of blood damage in bovine blood. All of these
results were submitted to the U.S. Food and Drug Administration and to
the auditing authorities in Europe prior to clinical use. The system was
tested on animals at the University Veterinary Clinic in Pisa, Italy,
and at the St. Gatien Clinic in Tours, France.
Mike van Driel, Dideco product manager and one of the engineers who helped
develop the concept, said, "The bovine testing was critical because
using a device that actively drains the patient directly is a big paradigm
shift for our market, and poses some significant training issues. We were
able to have a firm grasp on how to deal with these issues prior to initiation
of human clinical evaluations here in Europe."
The Synergy system is currently in clinical use in 11 centers in Europe
and at one in the United States, after having received premarket FDA approval
in April last year. More than 300 units have been used to date in the
European Union and the United States. The clinical use of the system in
Europe and the U.S. will continue to be expanded as the techniques for
managing the system are improved.
There are still some challenges to overcome in managing the system, including
accidental introduction of air at the surgical site and the removal of
such air from the system. Air occasionally enters the circuit due to excess
negative pressures or small leaks at the cannulation site of the heart,
and these air emboli must be removed.
In a traditional system, they are removed by the reservoir, while the
reservoirless Synergy device uses a large-capacity bubble trap before
the pump and an arterial filter after the pump to remove entrained air.
Air sensors on the tubing lines also alert the operator ahead of time,
and can be used in conjunction with a remote clamping system that will
prevent air from being passed to the patient.
Increased automation in the mini heart-lung machine is in the pipeline,
as is a second-generation device that is two-thirds the size of the current
system.
So while the doctors wait for clinically significant strides in the next
generation of minimally invasive cardiac surgery, Synergy represents another
potentially helpful tool for on-pump cardiac procedures.
This article was prepared by staff writers in collaboration
with outside contributors.
Return
to Index
|